- 1 side connected to solenoid valve connector without seal
("Plug and Play", as plug only has to be plugged onto measuring cell)- 1 side labelled cable with wire end ferruleswhite: PT100brown: PT100yellow: Measuring cell
SL green: Measuring cellblue: shielding
- 1 side connected to solenoid valve connector without seal
("Plug and Play", as plug only has to be plugged onto measuring cell)- 1 side labelled cable with wire end ferruleswhite: PT100brown: PT100yellow: Measuring cell
SL green: Measuring cellblue: shielding
Basic type:
ecoLine CR-PVC - Conductive 2-electrode conductivity sensor
|
Body material:
PVC
|
Cell constant:
C = 0.1 (measuring range 0.1 ... 1.0 mS/cm)
|
Cell material:
Stainless steel 1.4571
|
Electrical connection:
Device plug according to DIN EN 175301-803 (Hierschmann plug)
|
Measuring resistor (temperature sensor):
PT100
|
Process connection:
Screw-in thread G 3/4" A
2-electrode principleproven versions for industrial use measuring ranges 0 ... 1 mS/cm (K = 0.1) or 0.01 ... 15 mS/cm (K = 1.0) temperature range up to 55 °C, maximum pressure up to 6 bar at 25 °C JUMO ecoLine CR-PVC conductive 2-electrode conductivity sensorBrief description: Conductivity sensors of the 202923 series are used, for example, in general water technology, cooling and air-conditioning systems, drinking and bathing water, or in industrial rinsing and process water circuits. The body material PVC limits the application to medium temperatures up to 55 °C. Two versions with cell constants K = 0.1 or 1.0 are available. Versions with the common thread sizes G 3/4" or G 1" or a plug-in version are available for installation. The plug-in version is used, for example, with the matching PVC tees in nominal sizes DN 25, 32 and 40 and allows the sensor to be quickly dismantled for cleaning or maintenance. The electrical connection is made either via a detachable device plug or via a fixed cable connection. The overall construction and the parts in contact with the medium are physiologically harmless. The electrode material is stainless steel (K = 0.1) or special graphite (K = 1.0). The optionally integrable temperature sensor allows simultaneous recording of the medium temperature for temperature compensation in downstream measuring amplifiers.Stainless steel variant K = 0.1 The sensor has 3 metallic pin electrodes. The two outer pins are electrically connected and form one pole of the 2-electrode conductivity sensor. The counter electrode is the middle stainless steel pin. This design achieves very high measurement precision with low stray fields. A temperature sensor can be integrated in the middle pin. Graphite variant K = 1.0 For physical reasons, special graphite is used as the electrode material for higher conductance values. The two graphite electrodes are completely integrated in the electrode shaft. The optional temperature sensor is located in a stainless steel sleeve that is immersed in the measured medium. Operating principleThe sensors of the 202923 series are 2-electrode conductivity sensors. An AC voltage is applied to them by the transmitter. The current flowing across the liquid and the electrodes is determined by the conductivity of the liquid
LED Configuration:
Standard (see text)
|
measuring range:
0-2000 μS/cm
|
screw-in cel:
3/4"
High-quality conductivity meter with a 4-20mA signal output for forwarding the measured value to a higher-level control system, for example. The measurement is temperature compensated and therefore offers you an accurate measurement result. Also here we have the possibility to offer you an individual design regarding the connection, the LEDs including the measuring range up to a private label.
Due to the acrylic glass cover used, the light of the LEDs is widely scattered and the current LED status can be seen from a distance in darker rooms.
Quality Made in Germany.
Standard LED configuration (1 green LED, 1 red LED):
<1500 µS/cm = green
>1500 µS/cm to <1900µS/cm = green/red flashing
>1900 µS/cm = red
Technical data:
Measuring range: 0 - 2000 μS/cm
Temperature compensation: Linear 2.2 % / K, reference temperature 25°C
Accuracy conductivity: +/-10% of the maximum full scale value
Accuracy temperature: +-1.0 °C
Limit value display: Optical via LED
Analog output: 4-20 mA = 0-2000 μS/cm
Power supply: 24 VDC +-10%
Power consumption: < 2.5W
Screw-in thread: 3/4"
Housing: POM (polyoxymethylene)
Vision panel: PMMA (acrylic glass)
Electrodes: stainless steel 1.457
Dimensions: Diameter 60 mm x 60 mm (top) / 100 mm (total)
Gaskets electrodes: EPDM, 28 x 3 mm
Gasket housing: EPDM, 50 x 2 mm
Application limits: P max: 6 bar; T max: 50 °C
Protection class: IP 65*
LED Configuration:
Standard (see text)
|
measuring range:
0-50 μS/cm
|
screw-in cel:
Your desired thread
High-quality conductivity meter with a 4-20mA signal output for forwarding the measured value to a higher-level control system, for example. The measurement is temperature compensated and therefore offers you an accurate measurement result. Also here we have the possibility to offer you an individual design regarding the connection, the LEDs including the measuring range up to a private label.
Due to the acrylic glass cover used, the light of the LEDs is widely scattered and the current LED status can be seen from a distance in darker rooms.
Quality Made in Germany.
Standard LED configuration (1 green LED, 1 red LED):
<20 µS/cm = green
>20 µS/cm to <40µS/cm = green/red flashing
>40 µS/cm = red
Technical data:
Measuring range: 0 - 50 μS/cm
Temperature compensation: Linear 2.2 % / K, reference temperature 25°C
Accuracy conductivity: +/-10% of the maximum full scale value
Accuracy temperature: +-1.0 °C
Limit value display: Optical via LED
Analog output: 4-20 mA = 0-50 μS/cm
Power supply: 24 VDC +-10%
Power consumption: < 2.5W
Screw-in thread: according to your requirements
Housing: POM (polyoxymethylene)
Vision panel: PMMA (acrylic glass)
Electrodes: stainless steel 1.457
Dimensions: Diameter 60 mm x 60 mm (top) / 100 mm (total)
Gaskets electrodes: EPDM, 28 x 3 mm
Gasket housing: EPDM, 50 x 2 mm
Application limits: P max: 6 bar; T max: 50 °C
Protection class: IP 65*
LED Configuration:
Standard (see text)
|
measuring range:
Your desired measuring range
|
screw-in cel:
3/4"
High-quality conductivity meter with a 4-20mA signal output for forwarding the measured value to a higher-level control system, for example. The measurement is temperature compensated and therefore offers you an accurate measurement result. Also here we have the possibility to offer you an individual design regarding the connection, the LEDs including the measuring range up to a private label.
Due to the acrylic glass cover used, the light of the LEDs is widely scattered and the current LED status can be seen from a distance in darker rooms.
Quality Made in Germany.
Standard LED configuration (1 green LED, 1 red LED):
<20 µS/cm = green
>20 µS/cm to <40µS/cm = green/red flashing
>40 µS/cm = red
Technical data:
Measuring range: according your requirements
Temperature compensation: Linear 2.2 % / K, reference temperature 25°C
Accuracy conductivity: +/-10% of the maximum full scale value
Accuracy temperature: +-1.0 °C
Limit value display: Optical via LED
Analog output: 4-20 mA = 0-... μS/cm (according your needed measuring range)
Power supply: 24 VDC +-10%
Power consumption: < 2.5W
Screw-in thread: 3/4"
Housing: POM (polyoxymethylene)
Vision panel: PMMA (acrylic glass)
Electrodes: stainless steel 1.457
Dimensions: Diameter 60 mm x 60 mm (top) / 100 mm (total)
Gaskets electrodes: EPDM, 28 x 3 mm
Gasket housing: EPDM, 50 x 2 mm
Application limits: P max: 6 bar; T max: 50 °C
Protection class: IP 65*
Head transmitter:
4-20 mA
|
Tee:
1.5" flow cell with 1" inner thread NPT
|
measuring range:
0.1-1.000 μS/cm
The conductivity measuring package consists of a 3 meter connection cable, a 4-20mA head transmitter, and the conductivity sensor. Due to the design of the sensor assembly, only the sensor needs to be replaced, when it should be worn out . All other components of this system can be used continuously. For an optimal process integration the NeoTec sensor is delivered with a fitting T-piece.
Advantages:
preconfigured for NeoTecMaster®
modular design
industry standard
resource friendly
Head transmitter:
Modbus 12-24 VDC
|
Tee:
1.5" flow cell with 1" inner thread NPT
|
measuring range:
0.1-500 μS/cm
The conductivity sensor consists of a 3 meter connection cable, a Modbus head transmitter, and the conductivity sensor. Due to the design of the sensor assembly, only the sensor needs to be replaced. All other components of this system can be used continuously. For an optimal process integration the NeoTec sensor is delivered with a fitting T-piece.
Advantages:Preconfigured for NeoTecMaster®
Modular design
Industry standard
conserves resources
Cell constant:
1
|
Maximum temperature and pressure:
max. 100 °C / max. 13 bar (200 psig)
|
Measuring resistor (temperature sensor):
PT1000
|
Version:
1/2“ NPT
Our Neomeris Select high temperature conductivity cells are suitable in a wide range of high performance applications with high temperatures and pressures.To meet these demanding technical requirements, our sensors are manufactured from 316 stainless steel with PEEK elements and feature a one-piece cell design suitable for high pressure conditions. The conductivity cells are available with different cell constants (0.1/cm and 1.0/cm) and temperature sensors (PT100 and PT1000) and are designed for a specific pressure-temperature range. Certain model types also allow you to choose between a short and long thread, depending on the installation situation.Our 1/2" inch Select LF stainless steel conductivity cell is a general purpose sensor that can be used in a variety of standard applications such as in cooling towers or water treatment plants, etc. and is easy to install. In contrast, our Select HTLF Ultra high performance conductivity cell is the top of the range model. This conductivity sensor is specially designed for boiler applications where it is permanently integrated in the boiler. Via our selection menu you can easily choose the right conductivity sensor for your application. Please contact us if you have any questions.
Cell constant:
0.1
|
Maximum temperature and pressure:
max. 200 °C / max. 17 bar (250 psig)
|
Measuring resistor (temperature sensor):
PT1000
|
Version:
3/4“ NPT for installation in 3/4“ T-piece
Our Neomeris Select high temperature conductivity cells are suitable in a wide range of high performance applications with high temperatures and pressures.To meet these demanding technical requirements, our sensors are manufactured from 316 stainless steel with PEEK elements and feature a one-piece cell design suitable for high pressure conditions. The conductivity cells are available with different cell constants (0.1/cm and 1.0/cm) and temperature sensors (PT100 and PT1000) and are designed for a specific pressure-temperature range. Certain model types also allow you to choose between a short and long thread, depending on the installation situation.Our 1/2" inch Select LF stainless steel conductivity cell is a general purpose sensor that can be used in a variety of standard applications such as in cooling towers or water treatment plants, etc. and is easy to install. In contrast, our Select HTLF Ultra high performance conductivity cell is the top of the range model. This conductivity sensor is specially designed for boiler applications where it is permanently integrated in the boiler. Via our selection menu you can easily choose the right conductivity sensor for your application. Please contact us if you have any questions.
Cell constant:
1
|
Maximum temperature and pressure:
max. 252 °C / max. 41 bar (600 psi)
|
Measuring resistor (temperature sensor):
PT100
|
Version:
3/4“ NPT for installation in 3/4“ T-piece
Our Neomeris Select high temperature conductivity cells are suitable in a wide range of high performance applications with high temperatures and pressures.To meet these demanding technical requirements, our sensors are manufactured from 316 stainless steel with PEEK elements and feature a one-piece cell design suitable for high pressure conditions. The conductivity cells are available with different cell constants (0.1/cm and 1.0/cm) and temperature sensors (PT100 and PT1000) and are designed for a specific pressure-temperature range. Certain model types also allow you to choose between a short and long thread, depending on the installation situation.Our 1/2" inch Select LF stainless steel conductivity cell is a general purpose sensor that can be used in a variety of standard applications such as in cooling towers or water treatment plants, etc. and is easy to install. In contrast, our Select HTLF Ultra high performance conductivity cell is the top of the range model. This conductivity sensor is specially designed for boiler applications where it is permanently integrated in the boiler. Via our selection menu you can easily choose the right conductivity sensor for your application. Please contact us if you have any questions.
Analogue and digital communication for more process visibility• Modbus or 4-20 mA sensor output available• Configuration and calibration can be done using web based software
Corrosion, Coating and Fouling Resistant• Noryl body has a wide solvent tolerance and temperature stability to 105°C (used with flow cell 80°C, with integrated 4-20mA output or Modbus 70°C)
Mounting• Can be installed in line using the specific flow cell or submersed via the 3/4" MNPT thread
Wide Measurement Range• Can measure conductivity from 0 – 2.000.000 μS (0 - 2 S)